Miniaturized thin film glutamate and glutamine biosensors

   
Tyutrina G.V., Zharikova A.V., Zanine V.A., Berezov T.T., Mozer I., Jobst G., Aschauer E., Svasek P., Varahram M., Urban G.
PubMed Id: 9281221
Year: 1997 vol: 43  issue:1  pages: 20-28
Abstract: Integrated thin film biosensors for simultaneous measurement of L-glutamine and L-glutamate in a micro-flow cell were developed. Due to a novel glutaminase with an activity optimum in the neutral pH range a direct monitoring of glutamine in mammalian cell culture medium could be performed. The glutamine bienzyme sensor was prepared by coimmobilization of glutaminase with glutamate oxidase within a photopatterned poly(2-hydroxyethyl methacrylate) (pHEMA) hydrogel membrane. The sensor response is linear in the concentration range of 50 mmol to 10 mmol glutamine/l. Additionally a glutamate biosensor is integrated on the sensor chip for difference measurement of possible glutamate interferences. The sensorchip can be used for at least 300 measurements without any alteration in the performance of its sensors. A new sensorchip with an integrated flow cell provides the possibility of simultaneous measurement of four different parameters at a cell volumina of 1 m1. Completing the microsystem a battery operated surface mounted device (SMD) potentiostat was developed to get a 'lab on chip'.
Download PDF:
Reference: Tyutrina G.V., Zharikova A.V., Zanine V.A., Berezov T.T., Mozer I., Jobst G., Aschauer E., Svasek P., Varahram M., Urban G., Miniaturized thin film glutamate and glutamine biosensors, Voprosy meditsinskoi khimii, 1997, vol: 43(1), 20-28.
References
 2002(Vol:48)
 2001(Vol:47)
 2000(Vol:46)
 1999(Vol:45)
 1998(Vol:44)
 1997(Vol:43)
 1996(Vol:42)